Measuring
Piezoresistive transducers
MGCplus A0534-30.0 HBM: public 185
10. Use to select the
Adjust amplifier
button and confirm with .
11. Use the cursor keys
to select the
calibrate...calibrate...
button and confirm
with
.
12. Unload the transducer.
13. In the left edit field for characteristic curve point 1 enter the value 0 and
confirm with
.
14. Use
to select the
measure
button in line Point 1 and confirm with
.
15. Use
to select the left edit field in line Point 2, enter the value 250
and confirm with
.
16. Apply the test pressure to the transducer.
17. Use the cursor keys
to select the
measure
button in line Point 2. Now
if you press
, a measurement starts and the current measured value
appears in mv/V on the left next to the
measure
button.
18. Use
to select the
OK
button and confirm with (the amplifier
converts the nominal (rated) value to 300 bar. The calibration data for
250 bar remains intact).
19. Use
to go to measuring mode and confirm the confirmation prompt
with
.
AmplifierSystem Display Options
Transducer
Signal conditioning
Display
Analog outputs
Switch
bar
0.0000Point 1:
mV/V
0.0021
Shunt off
CancelOK
measure
250.000Point 2 166.67
measure
Characteristic curve points
Unit:
2.0000 ...
Type Full bridge circuit or high level
Current feed
Unit: mV/V
Zero pt.:
Nom. val.:
Gage factor:
TRANSDUCER CHANNEL1
bar
0.0000 ... 0.0000 ...
10.0000 ...
calibrate...calibrate... Adjust amplifier
0.0000 ...
5V
measure