Measuring
General principles for adjusting a measurement channel
MGCplus A0534-30.0 HBM: public 131
7.2.4 Display
Unit1
Select the unit in the display. You can choose between the basic unit (mV/
V), the user unit (kg, etc.) and the unit of the analog output (V). Then the
corresponding measured value appears in the display.
You can achieve the same effect with the "Unit" function (factory settings:
F3/level 2;
è see also section 9.2 "F keys", page 232).
Abs
If the absolute display is activated (
3
), the signal appears in the display at
the amplifier input without signal conditioning (for example zero drift or tar
ing).
Decimal places
Number of places after the decimal.
Display range
Display range in the selected unit (displayed automatically).
Step
The step defines the size of the display jumps. It refers to the last decimal
place of the nominal (rated) value.
Example: Nominal (rated) value 20kg
Decimal places 1 (20.0kg)
Step 1 means display jumps every 100g
Step 5 means display jumps every 500g
Decimal places 3 (20.000kg)
Step 1 means display jumps every 1g
Step 5 means display jumps every 5g
Unit Abs
Decimal places
Display range from
to
Step
DISPLAY CHANNEL1
100
10.000 V
−10.000 V
3 ...
V
3
Unit Abs
Decimal places
Display range from
to
Step
DISPLAY CHANNEL1
100
10.000 V
−10.000 V
3 ...
V
3
AmplifierSystem Display Options