EasyManua.ls Logo

HBM MGCplus - Electromagnetic Compatibility

HBM MGCplus
304 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electromagnetic compatibility
14 A0534-30.0 HBM: public MGCplus
The device complies with the safety requirements of DIN EN 61010 Part 1
(VDE 0411 Part 1); protection class I.
The insulation resistance of the connecting cables (v50V) must be at least
350V(AC).
1.1 Electromagnetic compatibility
The MGCplus device has been tested based on EMC product standard
EN 61326-1:2013. This standard includes definitions of limit values and test
levels for various electromagnetic environments.
Regarding emission (EME), requirements are included for class A (industrial
environments) and class B (residential, business and commercial environ
ments as well as small businesses). Laboratory applications also usually
require class B.
The product standard here references to EN 55011:2009+A1:2010.
Regarding immunity to interference, the product standard includes require
ments for controlled electro-magnetic environments (lowest requirements),
general environments and industrial environments (highest requirement).
MGCplus meet the following requirements:
-
Emission (EME): Class B
- Immunity to interference: Industrial environment
The MGCplus series and the individual modules thus essentially meet the
highest requirements and are therefore suitable for use in all environments
described in the product standard.

Table of Contents

Related product manuals