EasyManua.ls Logo

HBM MGCplus - Calibrating the Characteristic Curve of the Transducer

HBM MGCplus
304 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measuring
Adapting to the transducer
148 A0534-30.0 HBM: public MGCplus
7.3.3.2 Calibrating the characteristic curve of the transducer
By transferring the signals supplied by the transducer at a defined
deflection
Example: A displacement transducer with a nominal (rated) displacement of
20 mm (nominal (rated) value 10 mV/V) is set with a gaging block of 10 mm,
but the measuring range should be 15 mm.
Information
If the zero point and nominal (rated) sensitivity are not changed (for exam
ple in a recalibration), you can skip steps 1-12.
1. Move the transducer to the zero position.
2. Use
or a direct entry to choose the desired channel (for example 3.2
).
3. Use
to switch to Setup mode.
4. Press
.
5. In the pull-up menu select Transducer and confirm with
.
6. Use
to switch to the Current feed selection field, press and
select 2.5V.
7. Confirm with
.
8. Use
to switch to the Unit selection field and press . Select mm
as the unit and confirm with
.
9. Use
to switch to the Zero point edit field and enter the value 0 in
the left edit field. Confirm with
.
mm
0.0000Point 1:
mV/V
0.0021
Shunt off
CancelOK
measure
10.000Point 2 5.0018
measure
Characteristic curve points
Unit
Type IND half bridge circuit
Current feed
Unit: mm mV/V
Zero pt.:
Nom. val.:
Gage factor:
AmplifierSystem Display Options
Transducer
Signal conditioning
Display
Analog outputs
Switch
10.0000 ...
TRANSDUCER CHANNEL1
0.0000 ... 0.0000 ...
20.0000 ...
calibrate... Adjust amplifier
0.0000 ...
measure
2.5V

Table of Contents

Related product manuals