Measuring
Adapting to the transducer
164 A0534-30.0 HBM: public MGCplus
7.3.7 Thermocouples
Thermocouples are active transducers. To perform measurements with ther
mocouples you will need the ML801B amplifier module as well as connec
tion board AP809. The temperature cold junction is built into the AP809. The
amplifier module performs cold-spot compensation and linearization for the
thermocouple type K.
If you selected 'Thermocouples' as the operating mode and °C or °F as the
unit, the corresponding temperature display will be in the selected unit. If
you select
operating mode "75 mV", you will have the output voltage of
the thermocouple directly (without linearization and cold-spot
compensation).