Introduction
Conditions at the place of installation
MGCplus A0534-30.0 HBM: public 29
3.8 Conditions at the place of installation
CAUTION
S Protect the devices in a desktop housing from moisture and dampness
or weather conditions such as rain, snow, etc.
S Make sure that you do not cover the ventilation openings at the side, the
openings for the power pack fan on the back side of the device and the
openings underneath the device.
S Do not expose the device to direct sunlight.
S Comply with the maximum permissible ambient temperatures for the sys
tem devices, as stated in the technical data sheet.
S For installation in 19" electrical enclosures, due to poorer heat dissipa
tion, measures must be taken to ensure that the maximum permitted am
bient temperature (refer to the technical data sheet) is not exceeded! We
recommend forced venting in any case and in especially critical cases
intermediate spaces between the upper and lower rack frames.
S The devices are classified in overvoltage category II, degree of pollution 2.
S Install the device so that it can be disconnected from the mains at any
time without difficulty.
S It is safe to operate the MGCplus up to an altitude of 2000 m.