146
Saving Measurement Data
(3) PARA (measurement parameters)
Sets whether to use the save measurement parameters as the text le’s header.
1
Select either the ON key (save as the
header) or the OFF key (do not save as
the header).
2
Touch the EXIT key.
The dialog box will be closed.
When ON When OFF
"HIOKI E.E. CORPORATION","IM3536","Ver. 1.00",
"Serial No. 123456789"
"DATE","11-11-30"
"TIME","10:10:06"
"FREQ","1.0000E+03","Hz"
"V","1.000","V"
"LIMIT","OFF"
"RANGE","AUTO","10k","ohm"
"LOW Z","OFF"
"JUDGE SYNC","OFF"
"JUDGE","OFF"
"SPEED","MED"
"TRIG","INT"
"AVG","OFF"
"DELAY","0.0000","s"
"TRIG SYNC","OFF"
"DCBIAS","OFF"
"OPEN","OFF"
"SHORT","OFF"
"LOAD","OFF"
"CABLE","0","m"
"SCALE","OFF"
"Z[ohm]","OFF","PHASE[deg]","OFF"
"4.983329E+03","","0.074",""
"HIOKI E.E. CORPORATION","IM3536","Ver. 1.00",
"Serial No. 123456789"
"DATE","11-11-30"
"TIME","10:10:06"
"FREQ","1.0000E+03","Hz"
"V","1.000","V"
"LIMIT","OFF"
"RANGE","AUTO","10k","ohm"
"LOW Z","OFF"
"JUDGE SYNC","OFF"
"JUDGE","OFF"
"SPEED","MED"
"TRIG","INT"
"AVG","OFF"
"DELAY","0.0000","s"
"TRIG SYNC","OFF"
"DCBIAS","OFF"
"OPEN","OFF"
"SHORT","OFF"
"LOAD","OFF"
"CABLE","0","m"
"SCALE","OFF"
"4.983329E+03","","0.074",""
Display when ON:
No. 1 parameter: Z (impedance [
Ω
]), No. 2 parameter: OFF,
No. 3 parameter:
θ
(impedance phase angle [
°
]), No. 4 parameter: OFF