163
External Input/Output Connector and Signals
Pin I/O
*1
Signal name Function Logic
1 IN TRIG
――
External trigger
(See "Input (IN) signal function details" (p. 167).)
Rising/
falling
Edge
2 - (Unused) - - -
3 - (Unused) - - -
4 IN LD1
―――
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative Level
5 IN LD3
―――
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative Level
6 IN LD5
―――
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative Level
7 IN C1
――
Switches between the high-order digit and low-order digit
during BCD output.
Negative Level
8 - ISO_5V Isolated 5 V power output - -
9 - ISO_COM Isolated common signal ground - -
10 OUT ERR
―――
Outputs when a sampling error, overcurrent error, contact
error, Hihg-Z reject error, constant voltage/constant current
error, or voltage/current limit value exceeded error occurs.
Negative Level
11
*2
OUT BIN1
―――
Generates output when the BIN measurement result is BIN1.
Negative Level
PARA1-HI
――――
Generates output when the comparator judgment result is
HI for the No. 1 parameter.
D1-0
――
BCD output signal
12
*2
OUT BIN3
―――
Generates output when the BIN measurement result is BIN3.
Negative Level
PARA1-LO
―――― ――
Generates output when the comparator judgment result is
LO for the No. 1 parameter.
D1-2
――
BCD output signal
13
*2
OUT BIN5
―――
Generates output when the BIN measurement result is BIN5.
Negative Level
PARA3-IN
―――――
Generates output when the comparator judgment result is
IN for the No. 3 parameter.
D2-0
――
BCD output signal
14
*2
OUT BIN7
―――
Generates output when the BIN measurement result is BIN7.
Negative Level
AND
―――
Outputs judgment results for two parameter measured val-
ues (the result of a logical AND operation).
Generates output when both judgment results are IN. Ad-
ditionally, generates output when either of the No. 1 or No. 3
parameters is IN and the other is undetermined.
D2-2
――
BCD output signal
15 OUT BIN9
―――
Generates output when the BIN measurement result is BIN9.
Negative Level
D3-0
――
BCD output signal
16 OUT D3-2
――
BCD output signal Negative Level
17 OUT D4-0
――
BCD output signal Negative Level
18 OUT D4-2
――
BCD output signal- Negative Level
19 OUT OUT_OF_BINS
――
Generates output
when
the BIN judgment result does not match
any BIN.
Negative Level
20 IN C2
――
Switches between the No. 1 parameter and No. 3 parameter
during BCD output.
Negative Level
21 IN CALIB
――
DC adjustment request
(See "Input (IN) signal function details" (p. 167))
Negative Level
22 IN LD0
―――
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative Level
9
External Control