166
External Input/Output Connector and Signals
Pin I/O
*
Signal name Function Logic
19 - (Unused) - - -
20 - (Unused) - - -
21 - (Unused) - - -
22 - (Unused) - - -
23 - (Unused) - - -
24 - (Unused) - - -
25 - (Unused) - - -
26 - (Unused) - - -
27 - ISO_COM Isolated common signal ground - -
28 OUT EOM
――
This signal indicates that measurement is complete. “HI
(OFF)” indicates that measurement is in progress, while
“LO (ON)” indicates that measurement is complete. When
LO (ON), the comparator judgment results have been
nalized.
Falling Edge
29 OUT INDEX
――
Signal indicating that A/D conversion for the measurement
circuit has completed: When this signal changes from HI
(OFF) to LO (ON), the sample may be changed.
Falling Edge
30 OUT No.1_PARA1-IN
―――――
Outputs IN comparator judgment results for the rst No. 1
parameter.
Negative Level
31 OUT No.1_PARA3-HI
――――
Outputs HI comparator judgment results for the rst No. 3
parameter.
Negative Level
32 OUT No.1_PARA3-LO
―――― ――
Outputs LO comparator judgment results for the rst No. 3
parameter.
Negative Level
33 OUT No.2_PARA1-HI
――――
Outputs HI comparator judgment results for the second
No. 1 parameter.
Negative Level
34 OUT No.2_PARA1-LO
―――― ――
Outputs LO comparator judgment results for the second
No. 1 parameter.
Negative Level
35 OUT No.2_PARA3-IN
―――――
Outputs IN comparator judgment results for the second
No. 3 parameter.
Negative Level
36 - (Unused) - - -
37 - (Unused) - - -
*: IN indicates signal input to the instrument, while OUT indicates signal output from the instrument.