Measurement Range and Accuracy
Whether a given range is available varies with settings (cable length setting and DC bias setting).
[Cable length 0 m]
Range
Measurement frequency
DC
4 Hz to
99.99 Hz
100 Hz to
999.99 Hz
1 kHz to
10 kHz
10.001 kHz
to 100 kHz
100.01 kHz
to 1 MHz
1.0001 MHz
to 5 MHz
5.0001 MHz
to 8 MHz
100 M
Ω
10 M
Ω
Not available
1 M
Ω
100 k
Ω
Available
10 k
Ω
1 k
Ω
100
Ω
10
Ω
1
Ω
100 m
Ω
Not available
: Unavailable when DC bias is ON but may be set when DC bias is OFF.
[Cable length 1 m]
Range
Measurement frquency
DC
4 Hz to
99.99 Hz
100 Hz to
999.99 Hz
1 kHz to
10 kHz
10.001 kHz
to 100 kHz
100.01 kHz
to 1 MHz
1.0001 MHz
to 5 MHz
5.0001 MHz
to 8 MHz
100 M
Ω
10 M
Ω
Not available
1 M
Ω
100 k
Ω
10 k
Ω
Available
1 k
Ω
100
Ω
10
Ω
1
Ω
Not available
100 m
Ω
: Unavailable when DC bias is ON but may be set when DC bias is OFF.