EasyManua.ls Logo

Hioki IM3536 - Page 56

Hioki IM3536
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50
Setting Measurement Conditions (basic settings)
For setting range and accuracy
Open circuit voltage (V) mode and Constant voltage (CV) mode setting
Normal operation
When low-Z high-precision mode
(p. 55) is on
Open circuit voltage setting range 0.010 V to 5.000 V 0.010 V to 1.000 V
Open circuit voltage accuracy
1 MHz or less: ±10% rdg. ±10 mV,
1.0001 MHz or more: ±20% rdg. ±10 mV
Output impedance 100
Ω
±10
Ω
10
Ω
±2
Ω
For some samples, you may not be able to perform constant-voltage measurement (measurement in con-
stant-voltage mode) In this situation, the following error will be displayed:
Constant voltage measurement will not be performed. Change the constant-voltage value to a value that is
less than or equal to the value displayed for Vac.
(Example: Constant-voltage measurable range when measuring a C value of 1 µF at 10 kHz)
The sample impedance
Zm
is as follows:
Zm
=
Rm
+
jXm
=
0
[
Ω
] -
j15.9
[
Ω
]
Xm =
-1
(2πfC
)
The impedance
Zm’
as seen from the instrument’s voltage generator is as follows:
Zm’
=
Ro
+
Zm
=
100
[
Ω
] -
j15.9
[
Ω
] Ro:
Output resistance (100 [
Ω
])
Accordingly, the voltage
Vm
across both leads of the sample is as follows:
Vm =
|
Zm
|
×Vo
|
Zm’
|
=
15.9
[
Ω
Vo
101.3
[
Ω
]
V
o
:
generator output
Since the instrument’s voltage generator output range is 10 [mV] to 5 [V] as per the above table, the con-
stant-voltage measurable range is
Vm
= 1.6 [mV] to 0.78 [V] based on the above formula.
When low-Z high-precision mode is enabled, the output resistance
Ro
will be 10 [
Ω
].

Table of Contents

Related product manuals