57
Setting Measurement Conditions (basic settings)
Example: If the number of averaging iterations is set to 4 (measurement count, measured value output
points, and method for calculating measured values at output)
(1) Moving average
M1+M2+M3+M4
4
M4
M1+M2+M3
3
M3
M1+M2
2
M2
M1
1
M1
M5+M6+M7+M8
4
M8
M4+M5+M6+M7
4
M7
M3+M4+M5+M6
4
M6
M2+M3+M4+M5
4
M5
M6+M7+M8+M9
4
M9
M7+M8+M9+M10
4
M10
Measurement points
Measurement
value output
Elapsed time
(2) Arithmetic mean
M1+M2+M3+M4
4
M4M3M2M1
M5+M6+M7+M8
4
M8 M7 M6
M5
M9M10
Measurement points
Measurement
value output
Elapsed time
3