Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
87
Overview
Figure 44. Quality Control Configuration Screen—Manufacturer's
Controls/Calibrators Tab (Example)
Screen Buttons Description
1. Add Select this button to override the mean and
standard deviation values on the Levey-Jennings
Report for a manufacturer control/calibrator.
2. Edit Select a control/calibrator from the table and then
select this button to edit the mean and standard
deviation values for the chosen manufacturer
control/calibrator.
3. Delete Select a control/calibrator from the table and then
select this button to delete the override values for
the selected quality control/calibrator.
Table 49. Quality Control Configuration Screen Button Descriptions
(Manufacturer's Controls/Calibrators Tab)
Reflex Testing Configuration Screen
To access the Reflex Testing Configuration screen, select the Reflex Testing
Configuration button on the Administration screen (see Administration Screen on page
76).