Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
221
System Administration
2. Select the Warning During Sample Loading When Remaining Tests Are Below
field.
A keypad window opens.
3. From the keypad window, select the desired waste capacity level that will generate a
system alert.
4. Select OK when finished.
5. Shut down and restart the software by selecting the Instrument Shutdown button on
the Logoff/Shutdown screen when in the Setup or Ready state. Restart the software
from the Desktop Shield by selecting the Panther button.
Universal Fluid Management
Use the option available in the Universal Fluid Management expansion menu to
configure the Panther fluids warning and critical threshold.
Select and configure the following option within the menu:
Option Description
Panther
Fluid Icon
Color
Changes to
Yellow When
Remaining
Tests Are
Below
Select this field to set the lower limit warning alarm threshold
(between 0 to 600) for the Panther Universal Fluids inventory.
Set the number for the remaining tests that will trigger a
system alarm. The system will alert the operator by changing
the Panther Universal Fluids Icon to yellow when this warning
threshold has been met. The software must be restarted
before the selected configuration setting is enabled in the
software.
Additional instructions provided below.
Panther
Fluid Icon
Color
Changes to
Red When
Remaining
Tests Are
Below
Select this field to set the lower limit critical alarm threshold
(between 0 to 600) for the Panther Universal Fluids inventory.
Set the number for the remaining tests that will trigger a
system alarm. The system will alert the operator by changing
the Panther Universal Fluids Icon to red when this critical
threshold has been met. The software must be restarted
before the selected configuration setting is enabled in the
software.
Additional instructions provided below.
Table 97. Universal Fluid Management Options
From the expanded Universal Fluid Management menu on the System Configuration
screen, select the desired threshold.
To set the thresholds for Panther universal fluids: