23
Specifications and Requirements
μL, plus the dead space volume of 200 μL. See the appropriate Aptima package
insert for more information.
Tube/Specimen Type Dead Space Volume Required
Aptima Specimen Transfer Tube 1300 μL
Aptima Urine Specimen Transport
Tube Urine specimens
1300 μL or 700 µL*
Swab Specimen Transport Tube
Unisex swab specimens
1300 μL
Swab Specimen Transport Tube
Vaginal swab specimens
1300 μL
Cervical Specimen Collection and
Transport specimens
1300 μL
Multitest Swab Transport Tube 1300 μL
Primary Collection Tube Plasma or
serum
700 μL**
Specimen Aliquot Tube (SAT)
Plasma or serum
200 μL
*Dead space volume differs depending on assay. Refer to the appropriate assay
package insert for dead space volume requirements.
**1.2 mL above the pellet.
Table 12. Panther System Dead Space Volume Requirements
System Capacities and Throughput
Assay Reagent Kits Up to 4 assay reagent kits at a time
Universal Fluids Fluid Kit A and B 2,000 tests
Sample Racks
Up to 15 tubes per rack, up to 8 racks (120
tubes)
Table 13. System Capacities and Throughput