EasyManua.ls Logo

Hologic Panther System - Enable Non-Barcode Sample Loading; Enable Auto Sample ID Assignment; Enable Detection and Flagging of Repeat Barcodes;Reject All Repeat Test Orders

Hologic Panther System
425 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
225
System Administration
Option Description
Require Retainers
for Sample Racks
Check this cell to require retainers for sample racks. Rack
retainers are mandatory on Sample Racks that contain
samples with pierceable caps, regardless of this setting.
Enable Worklist ID
Suffix
Check this cell to enable the use of worklist ID suffix
functionality. When enabled, the system will assign
worklist IDs based upon the sample defaults settings (see
Configuring Additional Sample Defaults on page 119).
Additional information provided below.
ISBT128 Barcodes
Enabled
Check this cell to accept specimens labeled with ISBT128
barcodes.
Additional information provided below regarding barcode
stripping.
Table 100. Specimen Loading Options (Continued)
Enable Non-Barcode Sample Loading
Do not enable this setting when connected to an LIS or when running unlabeled EQCs.
To remove barcode scanning functionality during sample loading, check the Enable Non-
Barcode Sample Loading configuration. In this scenario, the system automatically
assigns an SID to specimens based upon the sample default settings (see Sample
Defaults on page 117).
Enable Auto Sample ID Assignment
Do not enable this setting when connected to an LIS or when running unlabeled EQCs.
To allow auto-generated SIDs to be assigned to samples with unreadable barcodes,
check the Enable Auto Sample ID Assignment for Unreadable Barcodes
configurations and leave the Enable Non-Barcode Sample Loading configuration
unchecked. In this scenario the system assigns an auto-generated SID to all specimens
with unreadable barcodes using the following naming convention:
AGSIDMMDDRRxxxx*
(*xxxx: counting up consecutively from 1 on each new date)
Enable Detection and Flagging of Repeat Barcodes/Reject All Repeat Test
Orders
Configure the system to alert the operator prior to processing duplicate test orders through
the System Configuration screen. Check the Enable Detection and Flagging of Repeat
Barcodes cell to configure the system to alert the operator when a specimen with a

Table of Contents