Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
283
Appendix B. Processing Flags
Results Processing Flags
Code Description Outcome
a Invalid kinetic profile.
The result calculation
for the flasher analyte
curve fit has failed.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
af Invalid kinetic profile.
The post-kinetic
adjustment value for the
flasher analyte is out of
range.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
afc The post-kinetic
adjustment value for the
flasher is greater than
the maximum cutoff
value for the assay.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
ag Invalid kinetic profile.
The post-kinetic
adjustment value for the
glower analyte RLU is
out of range for the
calibrator or control.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
agc The post-kinetic
adjustment value for the
glower analyte is
greater than the
maximum cutoff value
for the assay.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
arl Invalid kinetic profile.
The flasher curve R
squared value is less
than the minimum cutoff
value for the assay.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Table 117. Results Processing Flags