Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
54
Overview
Interactive Item Description
3. Event ID Specific error code for message.
4. Description Explanation of system event that occurred.
5. Clear Select this button to remove selected messages from
the Messages screen.
6. Clear All Select this button to remove all messages from the
Messages screen.
7. Silence Alarm Select this button to silence any audible alarms.
8. Remote
Diagnostics
Select this button to activate the PRO360° software.
This task should be used in conjunction with Hologic
Technical Support assistance as part of
troubleshooting system issues. Refer to the Remote
Diagnostics section for further information on how to
perform this task (see Troubleshooting on page 254).
9. Send Logs Select this button to export system log files. This task
should be used in conjunction with Hologic
Technical Support assistance as part of
troubleshooting system issues. Refer to the Send
Logs section for further information on how to perform
this task (see Send Logs on page 259).
10. Comments Select this button to access the Comments screen
and add or delete comments (see Comments Screen
on page 54).
11. Shutdown Select this button to shut down the system. The
Shutdown button is only available when the system
is safe to shut down.
Table 28. Message Screen Descriptions (Continued)
Comments Screen
Access the Comments screen by selecting the Comments button on the Messages
screen (see Messages Screen on page 53).
Use the Comments screen as a notepad to leave notes for later viewing. The comment
text must be between three (3) and 300 characters. Use the Date/Time, User Name, or
Comment column headers at the top of the table to sort the posted comments as needed.
Once a note is added, a comments icon will be displayed in the Status icons location on