Operator’s Manual
AW-17791-001 Rev. 002 (EN)
17
Specifications and Requirements
Current Draw Average: 700 W
Peak: 1400 W
l 100 VAC circuit draws 13 amps
l 240 VAC circuit draws 5.4 amps
Fuse Thermal circuit breaker
Table 6. Power Requirements (Continued)
Internal Specifications
Module Specifications
Incubators
Incubator 1: High Temperature
(HT)
63.5–64.5 °C
Incubator 2: Transition (TRAN)
43.2–44.7 °C
Incubator 3: Amplification (AMP)
42.4–43.0 °C
Load Stations
HPA
70.5–73.5 °C
AMP
44–46 °C
Bays
Reagent
16.5–19.5 °C
Sample
15.5–22.5 °C
Pumps
Vacuum Pump
4.0–12.0 inHg, target 7.0 inHg
Ramps
Chiller
17–19 °C
Table 7. Internal Specifications