274
Appendix B. Processing Flags
Code Description Outcome
FDF3 Auto Detect 1 pump
dispense failure, which
may be due to a
hardware or fluid
connection problem.
Sample is invalid.
Check all Auto Detect 1 connections. Contact
Hologic Technical Support.
FDF4 Auto Detect 2 pump
dispense failure, which
may be due to a
hardware or fluid
connection problem.
Sample is invalid.
Check all Auto Detect 2 connections. Contact
Hologic Technical Support.
FL1 Fluorometer internal
error.
MTU is invalid.
Contact Hologic Technical Support.
FL3 Fluorometer internal
error.
MTU is invalid.
Contact Hologic Technical Support.
FLSOR Fluorometer signal out
of range.
MTU is invalid.
Contact Hologic Technical Support.
FW# Firmware error detected.
A number with a
corresponding device
will follow the flag.
Contact Hologic Technical Support.
HB High background
reading in the
luminometer prior to
Auto Detect dispense,
which may be due to a
hardware or reagent
preparation problem.
Sample is invalid.
Contact Hologic Technical Support.
IDF Incubator door failure. Contact Hologic Technical Support.
IFF (1-3) Incubator 1-3 fan failure. Contact Hologic Technical Support.
IHF (1-3) Incubator 1-3 heater-foil
failure.
Contact Hologic Technical Support.
Table 116. Assay Processing Flags (Continued)