Appendix B. Processing Flags
Code Description Outcome
DIL1 Preconfigured dilution 1
used for final
concentration.
Informational message only. No action
necessary.
DIL2 Preconfigured dilution 2
used for final
concentration.
Informational message only. No action
necessary.
DIL3 Preconfigured dilution 3
used for final
concentration.
Informational message only. No action
necessary.
DIL4 Preconfigured dilution 4
used for final
concentration.
Informational message only. No action
necessary.
DSP Sample pipetting delay
error.
Sample not tested.
Load sample again to reschedule for testing.
DUO The replicate was tested
using an assay installed
as Development Use
Only.
Informational message only. No action
necessary.
EQC Sample is an External
Quality Control
Informational message only. No action
necessary.
FDF1 Oil pump dispense
failure, which may be
due to a hardware or
fluid connection
problem.
Sample is invalid.
Check all Oil connections. Contact Hologic
Technical Support.
FDF2 Wash solution pump
dispense failure, which
may be due to a
hardware or fluid
connection problem.
Sample is invalid.
Contact Hologic Technical Support.
Table 116. Assay Processing Flags (Continued)