EasyManua.ls Logo

Hologic Panther System - Power Requirements; Table 5. Environmental Output; Table 6. Power Requirements

Hologic Panther System
425 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
16
Specifications and Requirements
Altitude Operation: Altitude equal to or less than 2,000 m
(6,562 ft) above sea level
Storage: As required for travel
Ambient Temperature Operating: 15–30 °C
Storage: 5–45 °C
Transport: -20–70 °C
Relative Humidity Operating: 20–85% non-condensing
Storage: 10–90% non-condensing
Transport: 10–90% non-condensing
Pollution Degree 2
Installation Class II
Table 4. Environmental Requirements (Continued)
Heat Dissipation 550 watts (1878 BTU/hour) during steady state
Drain Output None: waste fluids drain to removable containers
Noise Output < 60 db (A)
Table 5. Environmental Output
Power Requirements
Voltage 100–240 ± 10% VAC
Frequency 50–60 Hz, 1800 VA single phase
Current Input Minimum of 15 amp circuit (dedicated)
20 amp circuit (dedicated if used with optional UPS)
Table 6. Power Requirements

Table of Contents

Related product manuals