Operator’s Manual
AW-17791-001 Rev. 002 (EN)
267
Troubleshooting
6. Dispose of the used filter according to laboratory procedures.
7. On the new Vacuum Trap Filter, turn the vacuum trap filter knobs clockwise in the
closed position to make sure that the Vacuum Trap Filter is secure.
8. Connect the tubing on the filter to the Coalescing bottle in the drawer.