Note—Use only Dow Corning High Vacuum Grease.
9. Apply the grease to the two male connectors for the Liquid Waste Container. Ensure
the grease is applied to the entire surface of the O-rings.
10. Apply a small amount of Dow Corning High Vacuum Grease to two fingertips.
11. Apply the grease to the inside of the Liquid Waste Bottle fittings.
12. Change gloves.
13. Place the Liquid Waste Container back in the Waste Drawer.
14. Close the Waste Drawer.