AW-17791-001 Rev. 002 (EN)
378
Appendix C. System Messages
Code Message Possible Cause Solution
406.001.00079 Failed to export raw
data, reason: {0}
Failed to export raw
data because there
was a problem with
USB drive or a
problem with
copying to USB
drive.
1. Retry export.
2. Check USB flash
drive for damage.
3. Replace USB flash
drive, if necessary.
4. If the problem
persists, collect
logs using the Send
Logs feature (see
Send Logs on page
259).
406.001.00080 Failed to export LIS
data, reason: {0}
Occurs when the
operator has
selected the Export
LIS Data button on
the Results screen
and there is error
with the export.
1. Retry export.
2. Check USB flash
drive for damage.
3. Replace USB flash
drive, if necessary.
4. If the problem
persists, collect
logs using the Send
Logs feature (see
Send Logs on page
259).
406.001.00081 Failed to print
report, reason: {0}
Report was not
printed.
1. Retry Print
command.
2. Check printer for
errors.
3. Collect logs using
the Send Logs
feature (see Send
Logs on page 259).
406.001.00082 Failed to create
report for viewing,
reason: {0}
Report was not
created.
1. Retry View
command.
2. Collect logs using
the Send Logs
feature (see Send
Logs on page 259).
Table 118. System Messages (Continued)