Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
397
Appendix C. System Messages
Code Message Possible Cause Solution
407.020.00001 TCR Carousel
configuration has
changed. Please
restart the system.
COP parameter 19
is different from that
of previous software
session. Most likely
caused after the
21st position of the
TCR mixer is
installed.
1. Shutdown and
restart.
2. If the problem
persists, contact
Hologic
Technical
Support.
407.020.00002 TCR Carousel
scanning has
failed. Please
correct error(s) and
close carousel door
to retry.
A barcode in the
TCR Carousel is
either not present,
misaligned or
damaged.
Contact Hologic
Technical Support.
Table 118. System Messages (Continued)