EasyManua.ls Logo

JUMO LOGOSCREEN 601 - Annex; Technical Data; Analog Inputs

JUMO LOGOSCREEN 601
280 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
265
11 Annex
11 Annex
11.1 Technical data
11.1.1 Analog inputs
General
Thermocouples
Number Max. 6 (see connection diagram)
A/D converter 24 bit delta-sigma
Sampling rate Up to 6 channels: 125 ms
Input filter Digital filter, 2nd order; filter constant can be set from 0 to 100.0 s
Galvanic isolation See "Galvanic isolation"
Designation Type Standard ITS Measuring range Accuracy
a
Fe-CuNi "L" DIN 43710 (1985) IPTS-68 -200 to +900 °C 0.1 %
Fe-CuNi "J" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 -210 to +1200 °C 0.1 % from -100 °C
Cu-CuNi "U" DIN 43710 (1985) IPTS-68 -200 to +600 °C 0.1 % from -100 °C
Cu-CuNi "T" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 -270 to +400 °C 0.1 % from -150 °C
NiCr-Ni "K" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 -270 to +1300 °C 0.1 % from -80 °C
NiCr-CuNi "E" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 -270 to +1000 °C 0.1 % from -80 °C
NiCrSi-NiSi "N" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 -270 to +1300 °C 0.1 % from -80 °C
Pt10Rh-Pt "S" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 -50 to +1768 °C 0.15 % from 100 °C
Pt13Rh-Pt "R" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 -50 to +1768 °C 0.15 % from 100 °C
Pt30Rh-Pt6Rh "B" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 0to182C 0.15 % from 600 °C
W5Re-W26Re "C" DIN EN 60584-1:2014
IEC 60584-1:2013
ITS-90 0to231C 0.1 % from 500 °C
W3Re-W25Re "D" ASTM E1751M-15 ITS-90 0 to 2315 °C 0.1 % from 500 °C
W5Re-W20Re "A1" GOST R 8.585-2001 ITS-90 0 to 2500 °C 0.1 % from 500 °C
Chromel®-Copel "L" GOST R 8.585-2001 ITS-90 -200 to +800 °C 0.1 % from -80°C
Chromel®-Alumel®
"K" GOST R 8.585-2001 ITS-90 -270 to +1372 °C 0.1 % from -80 °C
PLII (Platinel® II)
ASTM E1751M-15 ITS-90 0 to 1395 °C 0.1 %
Ambient temperature influ-
ence
100 ppm/K
Cold junction Internal (Pt100) or external (constant)
Cold junction accuracy (in-
ternal)
± 1K
Cold junction temperature
(external)
-30 to +85 °C (adjustable)
Base measuring range -20to+70mV
a
Accuracy refers to the measuring range

Table of Contents

Other manuals for JUMO LOGOSCREEN 601

Related product manuals