11.6 Sample racks
■■■■■■■■■■■■■■■■■■■■■■
106
■■■■■■■■
900 Touch Control
A maximum of 16 rack positions can be defined as special beakers for
each sample rack. The dialog Edit rack data / Special beakers shows a
list of all special beakers with their assigned rack positions.
[Edit]
Edit the data of the selected special beaker, see the following.
You can edit the data of the selected special beaker in the dialog Special
beaker / Edit.
Rack position
Number of rack position for selected special beaker. Any rack position can
be defined as a special beaker. It is, however, preferable to set them at
high rack positions in order to be able to begin sample series at rack posi-
tion 1. Rack positions defined as special beakers can no longer be used as
sample positions.
Input range 0 to 'maximum number of rack positions'