13 GLP manager
900 Touch Control
■■■■■■■■
133
[GLP test interval]
Defining the time interval for the GLP test (see "Dialogs "GLP test "Mea-
surement" / Test interval" and "GLP test "Titration" / Test interval"", page
133).
[Hardware]
Documenting the hardware with which the GLP test was carried out (see
"Dialogs "GLP test "Measurement" / Hardware" and "GLP test "Titra-
tion" / Hardware"", page 134).
Dialogs "GLP test "Measurement" / Test interval" and "GLP
test "Titration" / Test interval"
Test date
Date on which the GLP was carried out.
Format: YYYY:MM:DD
Monitoring
on | off (Default value: off)
If this parameter is activated, then the time interval after which a GLP test
has to be carried out again will be monitored.
GLP test interval
If you define a time interval for the GLP test, then the date in Next GLP
test will be tracked automatically.
Input range 1 to 999 days
Default value 999 days
Next GLP test
If you define a date for the next GLP test, then the GLP test interval will
be tracked automatically.
Format: YYYY:MM:DD
Action
Selection of the action which is carried out when the time interval has
expired.
Selection Display message | Document message | Cancel
determination
Default value Display message
For all three options it is documented in the deter-
mination data (see dialog More determination
data / Messages), that the time interval has been
expired.