29 Parameters
900 Touch Control
■■■■■■■■
279
More start conditions
Start meas. value
Before starting the titration, it will be dosed until this measured value is
reached. If the start measured value is reached by the dosing of a start
volume, then the titration starts directly.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
instrument used.
Measuring mode pH:
Input range –20.000 to 20.000
Selection off
Default value off
Measuring mode U, Ipol:
Input range
–2000.0 to 2000.0 mV
Selection off
Default value off
Measuring mode Upol:
Input range
–200.0 to 200.0 µA
Selection off
Default value off
Start slope
Before starting the titration, it will be dosed until this slope (measured
value per volume) is reached. If the start slope is reached by the dosing of
a start volume, then the titration starts directly.
Measuring mode pH:
Input range 00.000 to 9,999 pH/mL
Selection off
Default value off
Measuring mode U, Ipol:
Input range 0 to 999 mV/mL
Selection off
Default value off
Measuring mode Upol:
Input range
0 to 99 µA/mL
Selection off
Default value off