29.1 Dynamic equivalence point titrations (DET)
■■■■■■■■■■■■■■■■■■■■■■
280
■■■■■■■■
900 Touch Control
Dosing rate
Rate to be dosed with until the start measured value or the start slope is
reached.
Input range 0.01 to 166.00 mL/min
Default value 5.00 mL/min
Selection maximum
maximum
The maximum dosing rate depends on the cylinder volume (see chap-
ter 32, page 492).
Initial measured value
The [Initial meas. value] is measured before carrying out the start con-
ditions. This value is used for the titration direction determination,
which is necessary for the start measured value and stop measured
value parameters.
Signal drift
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.
Measuring mode pH, U and Ipol:
Input range 0.1 to 999.0 mV/min
Selection off
Default value off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Measuring mode Upol:
Input range
0.01 to 99.90 µA/min
Selection off
Default value off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.