29.9 Measurements (MEAS)
■■■■■■■■■■■■■■■■■■■■■■
380
■■■■■■■■
900 Touch Control
NOTICE
A constant measured value is often only reached after a certain time, as
mixing and the reaction itself require a certain time. The response time
of an electrode can also increase with time, i.e., reaching a constant
measured value takes longer and longer. Drift-controlled measured
value acceptance is particularly advisable in such cases, as the measured
values are only accepted when equilibrium has almost been reached.
Measuring mode pH, U, Ipol and Conc:
Input range
0.1 to 999.0 mV/min
Default value 10.0 mV/min
Selection off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Measuring mode Upol:
Input range
0.01 to 99.90 µA/min
Default value 10.00 µA/min
Selection off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Measuring mode T:
Input range
0.1 to 999.0 °C/min
Default value 0.5 °C/min
Selection off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Measuring mode Cond:
Input range
0.001 to 999.000 (mS/cm)/min
Selection off
Default value off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.