29 Parameters
900 Touch Control
■■■■■■■■
381
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.
Input range 0 to 999999 s
Default value 0 s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.
Input range 0 to 999999 s
The default value depends on the measuring mode.
Selection off
off
The measurement is continued endlessly.
Stop meas. value
The measurement is canceled when the specified measured value has
been reached since the start of the measurement.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
instrument used.
Measuring mode pH:
Input range –20.000 to 20.000
Selection off
Default value off
Measuring mode U, Ipol:
Input range
–2000.0 to 2000.0 mV
Selection off
Default value off
Measuring mode Upol:
Input range
–200.00 to 200.00 µA
Selection off
Default value off
Measuring mode T:
Input range
–20.0 to 150.0 °C