29.9 Measurements (MEAS)
■■■■■■■■■■■■■■■■■■■■■■
382
■■■■■■■■
900 Touch Control
Selection off
Default value off
Measuring mode Conc:
Input range 0.001 to 999999.999
The unit corresponds to the one of the standards
with which the calibration of the sensor has been
carried out.
Selection off
Default value off
Measuring mode Cond:
Input range
0.0001 to 999.9999 mS/cm
Selection off
Default value off
Temperature
This parameter is not available with the MEAS T (temperature measure-
ment) and MEAS Cond (conductivity measurement) commands.
Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement = automatic or continuous is defined (see sensor dialog
of the command). This value is used for temperature correction in pH and
concentration measurements.
Input range –20.0 to 150.0 °C
Default value 25.0 °C
Time interval MP
Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.
Input range 0.1 to 999999.0 s
Default value 2.0 s