Trainee Guide TR-L2-M2000
Level 2 Maintenance Training – M2000 Dispensing Systems 4-2-20 P/N 73-0018-00 (Revision B)
Information Sheet 4-6-1
Symptoms and Causes of Scale Failure
Symptom Possible Cause Recovery
Power Connector - Connect Power Cable
Blank Display Screen
Scale Display - Call Service Technician
Vibration
- Check surrounding area for any
equipment that may be causing
vibration.
- If problem cannot be resolved,
call Service Technician.
Filter Settings - Call Service Technician
Scale Will Not Zero
Air Currents
- Check surrounding area for any
air currents.
- If problem cannot be resolved,
call Service Technician.
Scale Calibration - Recalibrate Scale
Air Currents
- Check surrounding area for any
air currents.
- If problem cannot be resolved,
call Service Technician.
Inaccurate
Measurements
Scale Display - Reconfigure and Zero Scale
Wrong Units Displayed Scale Configuration - Reconfigure scale
Overload Limits Exceeded
- Replace weigh station cup and
zero scale.