Command reference
R&S
®
ZNL/ZNLE
1004User Manual 1178.5966.02 ─ 20
Example: Starting with a 4-port analyzer's default logical port assignment
L1<-->P1, L2<-->P2, L3<-->P3, L4<-->P4,
the remote command
SOUR:LPOR1 1,2
by default (ALIGn=ON) generates aligned logical ports
L1<-->P1&P2, L2<-->P3, L3<-->P4.
With ALIGn=OFF, the result is
L1<-->P1&P2, L3<-->P3, L4<-->P4
and L2 does not exist.
SYSTem:COMMunicate:NET:HOSTname <HostName>
Sets or gets the host name of the instrument.
Parameters:
<HostName> Host name
*RST: A preset of the instrument doesn't change the host
name.
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess <Address>
Selects one of the USB-connected calibration units for calibration (see commands
SENSe<Ch>:CORRection:COLLect:AUTO...). This command is not necessary if
only one cal unit is connected.
Parameters:
<Address> Name (USB address) of a connected calibration unit (string vari-
able). The names of all connected cal units can be queried using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?.
Example:
SYST:COMM:RDEV:AKAL:ADDR:ALL?
Query the names of all connected calibration units.
SYST:COMM:RDEV:AKAL:ADDR 'ZV-Z52::1234'
Select the cal unit named 'ZV-Z52::1234' for calibration.
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port TOSM calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Manual operation: See "Cal Unit" on page 476
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
Queries the names (USB addresses) of all connected calibration units.
Example:
See SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Usage: Query only
Manual operation: See "Cal Unit" on page 476