Command reference
R&S
®
ZNL/ZNLE
1028User Manual 1178.5966.02 ─ 20
Setting parameters: 
<UnitRef> S | SINV | Y | Z | YREL | ZREL
Keyword for the physical unit of the response values. The
parameters form four groups: S and SINV select relative units
(dB) for the limit line. Y selects admittance units (S/Siemens). Z
selects impedance units (Ω). YREL and ZREL select dimension-
less numbers (U).
Usage:  Setting only
CALCulate<Chn>:LIMit:RDOMain:FORMat 
<UnitRef>
Deletes the existing limit line and (re-)defines the physical units of the response values
of the limit line. The units of the stimulus values are defined via CALCulate<Chn>:
LIMit:CONTrol:DOMain.
Tip: This command is complemented by CALCulate<Chn>:LIMit:RDOMain:
COMPlex and CALCulate<Chn>:LIMit:RDOMain:SPACing.
Suffix: 
<Chn>
.
Channel number used to identify the active trace.
Setting parameters: 
<UnitRef> COMPlex | MAGNitude | PHASe | REAL | IMAGinary | SWR |
GDELay | L | C
Keyword for the physical unit of the response values. The
parameters form four groups: COMPlex, REAL, IMAGinary, and
SWR select dimensionless numbers (U) for the limit line. MAG-
Nitude selects relative units (dB). PHASe selects phase units
(deg). GDELay selects time units (s). L selects inductance units
(H/Henry). C selects capacitance units (F/Farad).
Usage:  Setting only
CALCulate<Chn>:LIMit:RDOMain:SPACing <Format>
Deletes the existing limit line and (re-)defines the physical units of the response values
of the limit line. The units of the stimulus values are defined via CALCulate<Chn>:
LIMit:CONTrol:DOMain.
Tip: This command is complemented by CALCulate<Chn>:LIMit:RDOMain:
COMPlex and CALCulate<Chn>:LIMit:RDOMain:FORMat.
Suffix: 
<Chn>
.
Channel number used to identify the active trace.
Setting parameters: 
<Format> LINear | LOGarithmic | DB | SIC
Keyword for the physical unit of the response values.
Default unit: dB (irrespective of the parameter selected)
Usage:  Setting only
VNA command reference