EasyManua.ls Logo

R&S ZNL Series - Page 342

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VNA GUI reference
R&S
®
ZNL/ZNLE
342User Manual 1178.5966.02 ─ 20
Figure 10-4: Example of a multiple trigger configuration
The table in the "Trigger Manager" dialog contains several editable (white) or non-edit-
able (gray) columns. All settings are analogous to the general trigger settings in the
Trigger tab. Refer to the following sections:
"Sequence" on page 340
"FreeRun / External / Manual / Multiple Triggers" on page 339
"Signal Type" on page 341
"Delay" on page 341
The column "Slope / Level 2" is only available, if the instrument is equipped with a sec-
ond external trigger input. This second external trigger input is provided by the Aux.
Port connector, which is part of option R&S FPL1-B5 "Additional Interfaces" (R&S ZNL
only).
... /Source/ ...
The table defines all settings related to "Multiple Triggers" (see Chapter 10.5.3, "Trig-
ger tab", on page 338). For all measurement sequences the following trigger sources
are available:
"Free Run" means that a trigger signal is not required
"External 1" is the external trigger signal fed in at either the Trigger In connector on
the rear panel or pin 2 of the Aux. Port connector.
"External 2" is the external trigger signal fed in at pin 25 of the Aux. Port connector
on the rear panel.
"External 1 and 2"/"External 1 or 2" means that the measurement sequence is initi-
ated after the analyzer has received an event from external trigger signal 1 and/or
2. A Signal Type must be specified for both trigger signals.
Manual means that the trigger event is generated manually, by selecting the Man-
ual Trigger button.
Sweep Softtool

Table of Contents

Related product manuals