Command reference
R&S
®
ZNL/ZNLE
578User Manual 1178.5966.02 ─ 20
Example:
MMEMory:COMMent "ACP measurement with Standard
Tetra from 23.05."
MMEMory::MMEMory:STORe1:STATe 1, "ACP_T"
As a result, in the selection list for recall settings, the comment
"ACP measurement with Standard Tetra from
23.05." is added to the ACP entry.
Manual operation: See "Comment" on page 100
MMEMory:COPY <FileName>, <FileName>
This command copies one or more files to another directory.
Parameters:
<FileName> String containing the path and file name of the source file.
<FileName> String containing the path and name of the target file.
The path may be relative or absolute.
MMEMory:DATA
<FileName>[, <Data>]
MMEMory:DATA? <FileName>
This command writes block data into a file. The delimiter must be set to EOI to obtain
error-free data transfer.
When you query the contents of a file, you can save them in a file on the remote con-
trol computer.
The command is useful for reading stored settings files or trace data from the instru-
ment or for transferring them to the instrument
Parameters:
<Data> <block_data>
Data block with the following structure.
#
Hash sign.
<number>
Length of the length information.
<number>
Length information of the binary data (number of bytes).
<data>
Binary data with the indicated <number> of bytes.
Parameters for setting and query:
<FileName>
Common instrument commands