Command reference
R&S
®
ZNL/ZNLE
616User Manual 1178.5966.02 ─ 20
SYSTem:IDENtify[:STRing] <String>
This command defines the response to *IDN?.
Parameters:
<String> String containing the description of the instrument.
Manual operation: See "Identification String" on page 157
SYSTem:KLOCk <State>
This command locks or unlocks manual operation.
Parameters:
<State> ON
LLO (local lockout). The instrument can only be operated
remotely, not locally.
OFF
Unlocks the manual operation mode. To operate the instrument
locally again, you must execute SYST:COMM:INT:REM OFF or
select the "Local" softkey first.
*RST: state not affected by *RST
Example:
SYST:KLOC ON
Activates LLO (remote control only)
Example:
SYST:KLOC OFF
SYST:COMM:INT:REM OFF
You can operate the instrument locally.
SYSTem:LANGuage <Language>
This function is used to emulate previous R&S signal and spectrum analyzers.
Parameters:
<Language> "FSV" | "SCPI"
*RST: SCPI
Example:
SYST:LANG FSV
The commands and settings for the R&S FSV instrument are
used.
Manual operation: See "Language" on page 159
SYSTem:LXI:INFO
This command queries the "LXI" settings.
Return values:
<LXIInfo> <current version> | <LXI class> | <Computername> |
<MAC adress> | <IP adress> | <Auto MDIX>
String containing the current "LXI" parameters.
Common instrument commands