SDATa
Output as unformatted trace data; see CALCulate<Chn>:DATA
. Query only.
FSIData
Output or input at "Fixture Simulation Input" data access point.
<Data> Trace data either in ASCII or block data format, depending on
the current FORMat[:DATA] setting.
The column order must match the one used during export.
Example: Suppose that a TOSM calibration for ports 1 and 2 is active in
channel no. 1.
CALCulate:DATA:CALL:CATalog?
Query the traces available for CALCulate<Ch>:DATA:CALL?.
The response is 'S11,S12,S21,S22'.
CALCulate:DATA:CALL? SDATa
Return the complex response values of all traces. The traces in
the catalog list are read one after another: The response array
contains n (number of points) pairs of real and imaginary values
for S
11
, followed by n pairs of values for S
12
, S
21
, and S
22
.
CALCulate<Ch>:DATA:CALL:CATalog?
Returns all traces which are available for CALCulate<Ch>:DATA:CALL in channel
no. <Ch>. The response is a string parameter with all S-parameter traces in the current
channel or in the active system error correction; see example.
Suffix:
<Ch>
.
Channel number
Example: See CALCulate<Ch>:DATA:CALL
Usage: Query only
CALCulate<Ch>:DATA:CHANnel:ALL? <Format>
Reads the current response values of all traces of the selected channel.
Suffix:
<Ch>
.
Channel number
Query parameters:
<Format> FDATa | SDATa | MDATa
Output format for the S-parameter data, see
CALCulate<Chn>:DATA.
Return values:
<Data> Trace data either in ASCII or block data format, depending on
the current FORMat[:DATA] setting.
Usage: Query only