Command reference
R&S
®
ZNL/ZNLE
686User Manual 1178.5966.02 ─ 20
Parameters:
<OutFormat> MLINear | MLOGarithmic | PHASe | POLar | GDELay | REAL |
IMAGinary | SWR | LINPhase | LOGPhase | IMPedance |
ADMittance | DEFault | COMPlex | MDB | MLPHase |
MDPHase | MIMPedance | PIMPedance | PADMittance |
MADMittance | MPIMpedance | MPADmittance | INDX
DEFault means the default marker format is dynamically adjus-
ted to the selected trace format (CALCulate<Chn>:FORMat).
For the other marker formats see the table below and the
description in "Marker format" on page 199.
*RST: DEFault
Manual operation: See "Dflt Marker Frmt" on page 421
SCPI GUI
MLINear "Lin Mag"
MLOGarithmic
MDB (for R&S ZVR compatibility)
"dB Mag"
PHASe "Phase"
POLar
COMPlex (for R&S ZVR compatibility)
"Real Imag"
GDELay "Delay"
REAL "Real"
IMAGinary "Imag"
SWR "SWR"
LINPhase
MLPhase (for R&S ZVR compatibility)
"Lin Mag Phas"e
LOGPhase
MDPhase (for R&S ZVR compatibility)
"dB Mag Phase"
IMPedance "R + j X" for reflection measurement traces
"R + j X series" for transmission measurement
traces
PIMPedance "R + j X", for reflection measurement traces
"R + j X parallel" for transmission measurement
traces
ADMittance "G + j B" for reflection measurement traces
"G + j B series" for transmission measurement
traces
PADMittance "G + j B" for reflection measurement traces
"G + j B parallel" for transmission measurement
traces
VNA command reference