Command reference
R&S
®
ZNL/ZNLE
784User Manual 1178.5966.02 ─ 20
In the query form, the <Conductance> parameter must be omitted. The command
returns the conductance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port pair (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine)
<1|2|3> Index i of the conductance G<i> in the related lumped circuit
model. If unspecified the numeric suffix is set to 1.
Parameters:
<Conductance> Conductance G<i> for the specified circuit model.
Range: -1kS to 1 kS.
Increment: 1 pS (1E-12 S)
*RST: 0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<arg0> STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Circuit model whose conductance G<i> shall be set, see
Table 11-10
Manual operation: See "Network" on page 553
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>[:STATe]
<Boolean>
Enables or disables the embedding function for port set <ListId>. It is allowed to
change the embedding network while embedding is enabled.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port set (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:EMBedding<ListId>:
DEFine)
Parameters:
<Boolean> OFF (0): Embedding inactive
ON (1): Embedding active
*RST: OFF (0)
VNA command reference