Command reference
R&S
®
ZNL/ZNLE
787User Manual 1178.5966.02 ─ 20
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:DATA <Interchange>, <arg1>
Defines an embedding network for a single-ended port based on the given S-Parame-
ter traces.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
TNDefinition to select the adequate circuit model before executing this com-
mand.
●
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt> to load cir-
cuit data from a Touchstone file located at the R&S ZNL/ZNLE's file system.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Setting parameters:
<Interchange> FPORts | IPORts | SGATes
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
<arg1> <block_data>
Content of a Touchstone file (*.s2p) in IEEE488.2 Block data
format.
Usage: Setting only
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp> <CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>? <CircuitModel>
Specifies the conductance value G in the "shunt L, shunt C" circuit model for single
ended port deembedding.
In the query form, the <Conductance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
<Cmp> Number of the conductance component in the circuit model. The
total number of conductances depends on the selected circuit
model.
VNA command reference