Example:
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:L1? SLL
Query the default inductance L1 for the Shunt L, serial L circuit
model. The response is 1E-009 (1 nH).
CALC:TRAN:VNET:SEND:DEEM:PAR:L1 SLL, 2.2E-9
Increase the inductance to 2.2 nH.
Manual operation: See "Network" on page 551
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:R<Cmp> <CircuitModel>, <Resistance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:R<Cmp>? <CircuitModel>
Specifies the resistance value R<Cmp> in the different circuit models for single ended
port deembedding.
In the query form, the <Resistance> parameter must be omitted. The command returns
the resistance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
<Cmp> Number of resistance in circuit model. The total number of resis-
tances depends on the selected circuit model.
Parameters:
<Resistance> Resistance R<Cmp> for the specified circuit model.
Range: -10 MΩ to 10 MΩ.
Increment: 1 mΩ (1E-3 Ω)
*RST: 0 Ω for all resistances connected in series with an
inductance. 10 MΩ for all resistances connected in
parallel with a capacitance
Default unit: Ohm
Parameters for setting and query:
<CircuitModel> CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL | SHLC | GSL |
LSG | SLG | SGL
Possible circuit models (character data); see Table 11-9.
Example:
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:R1? CSL;
R2? CSL
Query the default resistances for the Serial C, shunt L circuit
model. The response is 10000000; 0.
CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3
Increase the resistance R2 to 2.2 kΩ.
Manual operation: See "Network" on page 551