EasyManua.ls Logo

R&S ZNL Series - Page 799

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
799User Manual 1178.5966.02 ─ 20
ISMith Displays z in an inverted Smith diagram
GDELay For frequency sweeps only
Calculates the group delay at the related sweep
point and displays it in a Cartesian diagram
REAL Calculates Re(z) = x and displays it in a Cartesian
diagram
IMAGinary Calculates Im(z) = y and displays it in a Cartesian
diagram
SWR Calculates the standing wave ratio (1 + |z|) / (1 - |z|)
and displays it in a Cartesian diagram
LOGarithmic Displays z in a Cartesian diagram with logarithmic
scale
CALCulate<Chn>:FORMat:WQUType <Unit>
Selects the physical unit of the displayed trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Unit> POWer | VOLTage
Power or voltage units
*RST: POWer
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'b1'
Create channel 4 and a trace named Ch4Tr1 to measure the
wave quantity b
1
. The trace becomes the active trace in channel
4.
CALC4:FORM:WQUT VOLT
Select voltage units for the created trace (identified by the suffix
4).
Manual operation: See "Show as" on page 407
CALCulate<Chn>:GDAPerture:SCOunt <Steps>
Defines an aperture for the calculation of the group delay as an integer number of fre-
quency sweep steps.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Steps> Number of steps
Range: 1 to 10000
*RST: 10
VNA command reference

Table of Contents

Related product manuals