Command reference
R&S
®
ZNL/ZNLE
800User Manual 1178.5966.02 ─ 20
Example:
*RST; :CALC:FORM GDEL
Select group delay calculation for the active trace.
CALC:GDAP:SCO 15
Select an aperture of 15 steps.
Manual operation: See "Aperture Points" on page 420
CALCulate<Chn>:IAVerage:MODE <Mode>
Selects the quantities to be averaged if infinite averaging is enabled
(CALCulate<Chn>:IAVerage[:STATe] ON.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Mode> MAGPhase | REIMag
Selects the averaging mode.
MAGPhase
Averaging of magnitude and phase (default) of the complex
trace value
REIMag
Averaging of real and imaginary part of the complex trace value
Manual operation: See "Mode" on page 386
CALCulate<Chn>:IAVerage[:STATe] <Boolean>
Turns infinite averaging on or off.
The averaging mode can be selected using CALCulate<Chn>:IAVerage:MODE.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF | 1 | 0
*RST: 0
Manual operation: See "Infinite Average/Restart" on page 386
CALCulate<Chn>:PHOLd <HoldFunc>
Enables, disables, or restarts the max hold and the min hold functions.
Suffix:
<Chn>
.
Channel number used to identify the active trace
VNA command reference

Table of Contents