Command reference
R&S
®
ZNL/ZNLE
825User Manual 1178.5966.02 ─ 20
Parameters:
<TraceName> DALL – all data traces
MALL – all memory traces
<string> – single trace identified by its trace name (string param-
eter), e.g. 'Trc4'.
<Boolean> ON | OFF – display or hide traces.
Example: 
*RST; :DISP:TRAC:SHOW? 'Trc1'
Reset the analyzer, creating the default trace 'Trc1'. The trace is
displayed; the query returns 1.
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON; :DISP:WIND2:TRAC:FEED
'CH4TR1'
Create diagram area no. 2 and display the generated trace in
the diagram area.
DISP:TRAC:SHOW DALL, OFF
Hide both traces in both diagrams.
DISP:TRAC:SHOW? DALL
Query whether all data traces are displayed. The response 0
means that at least one trace is hidden.
Manual operation: 
See "Table Area" on page 349
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:X:OFFSet <StimulusOffset>
Shifts the trace <WndTr> in horizontal direction, leaving the positions of all markers
unchanged.
Suffix: 
<Wnd>
.
Number of an existing diagram area (defined by means of
DISPlay[:WINDow<Wnd>][:STATe] ON).
<WndTr> Existing trace number, assigned by means of DISPlay[:
WINDow<Wnd>]:TRACe<WndTr>:FEED.
Parameters:
<StimulusOffset> Stimulus offset value.
The range and unit depends on the sweep type.
Default unit: NN
Example: 
*RST; :DISP:WIND:TRAC:X:OFFS 1MHZ; :DISP:WIND:
TRAC:Y:OFFS 10
Create the default trace and shift it horizontally by 1 MHz, verti-
cally by 10 dB.
Manual operation:  See "Stimulus" on page 384
VNA command reference