Command reference
R&S
®
ZNL/ZNLE
882User Manual 1178.5966.02 ─ 20
REDuce
Cumulative moving averages of the real and imaginary parts of
each measurement result, provides the most effective noise sup-
pression for the "Real" and "Imag" formats and for complex trace
formats.
MOVing
Simple moving averages of the real and imaginary parts of each
measurement result; similar to REDuce, but with finite history.
Manual operation: See "Mode" on page 324
[SENSe:]BANDwidth...
The [SENSe:]BANDwidth... commands set the bandwidth of the IF filter (measure-
ment bandwidth).The forms BANDwidth and BWIDth are equivalent.
[SENSe<Ch>:]BANDwidth[:RESolution]............................................................................882
[SENSe<Ch>:]BWIDth[:RESolution].................................................................................882
[SENSe<Ch>:]BANDwidth[:RESolution]:SELect................................................................ 883
[SENSe<Ch>:]BWIDth[:RESolution]:SELect......................................................................883
[SENSe<Ch>:]BANDwidth[:RESolution] <ResBandw>
[SENSe<Ch>:]BWIDth[:RESolution] <IF Bandwidth>
Defines the IF bandwidth of the analyzer (measurement bandwidth). Values between
1 Hz and 500 kHz can be set.
Suffix:
<Ch>
.
Channel number
Parameters:
<IF Bandwidth> Bandwidths can be set in 1 – 1.5 – 2 – 3 – 5 – 7 steps. The ana-
lyzer rounds up any entered value between these steps and
rounds down values exceeding the maximum bandwidth.
Range: See above
*RST: 10 kHz
Default unit: Hz
Example:
BAND 1.1
Set a IF bandwidth of approx. 1.1 Hz for channel 1.
BAND?
The analyzer returns the rounded bandwidth of 1.5 Hz.
Manual operation: See "Bandwidth" on page 322
VNA command reference