Command reference
R&S
®
ZNL/ZNLE
885User Manual 1178.5966.02 ─ 20
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Define the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
CORR:CKIT:FOP? 'N 50 Ohm'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in the active cal kit.
CORR:CKIT:FOP? 'N 50 Ohm','New Kit 1'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
Manual operation: See "One Port Standards / Two Port Standards" on page 501
Table 11-14: Set command parameters
One port standards Two port standards
Open (Offset)
Short
Match Reflect Sliding
match
Symmetric
network
Through,
Line
Attenuation
<Con-
nType>
mandatory
<CalKit-
Name>
mandatory (and must not be an empty string)
<Standar-
dLabel>
mandatory (but can be an empty string)
<MinFreq>,
<MaxFreq>
mandatory
<Delay-
Param>,
<Loss>,
<Z0>
mandatory mandatory mandatory mandatory not used mandatory not used
<C0>,
<C1>,
<C2>, <C3>
mandatory mandatory* mandatory* mandatory** mandatory** not used
<L0>, <L1>,
<L2>, <L3>
mandatory* mandatory mandatory* mandatory** mandatory**
OPEN |
SHORt |
MATCh |
<Resist-
ance>
optional*** optional*** optional*** mandatory:
OPEN |
SHORt |
<Resist-
ance>
mandatory:
OPEN |
SHORt |
<Resist-
ance>
<Port1>[,
<Port2>]
optional: <Port 1> optional: <Port 1>,<Port 2>
* values are ignored during calibration
** if OPEN is selected the residual inductance (L0,...,L3) are during calibration; if SHORt is selected the fringing capacitance
(C0,...,C3) are during calibration
*** must be provided if the definition is restricted to a single port. Values are ignored during calibration
VNA command reference